    YSTE-160V Table Type Veterinary Portable Chemistry Analyzer





    Product parameter

     

    Model YSTE160V
    Analytical System Fully automatic, Random Access
    Method End point, Rate, Fixed time, Immunoturbidimetric, 1-2 regent, Multistandard, reagent/serum blank, etc
    Assay Item 30 assay
    Throughput 1 Reagent 160 tests/hr
    2 Reagent 100 tests/hr
    Test Menu T.BILI         r-GT/GGT    CHO    Ca      APOB-1/B   CHE
    D.BILI        ALP/AKP     TG      CL      APOB       CREA
    TP           UREA        CK      P       CK-MB      Mg
    ALB          CREA        LDH    C        ASO        AFU
    Lactic Acid    ALT/GPT      UA    α-HBDH  CO2        FMN
    LP(a)         AST/GOT     GLU   AMY      LDL-C      HDL-C
    CHE          LA           TBA    ADA      HS-CRP    PA
    Fe            Cu           Zn     C3       C4          IgG
    IgM           IgA           and drug, toxicity,etc.
    Sample Position 30
    Including standard, control and STAT; either serum cup or primary tube available
    Sample Volume 1~100ul, 0.1ul/step
    Reagent Position 40
    Reagent Volume R1: 1~400ul, 0.1ul/step, R2: 1~400ul, 0.1ul/step
    Reagent/Sample Probe With liquid level sensor, collision sensor, Teflon coating and automatic washing station
    Reaction Time 0~999 seconds
    Cuvette High quality uv-transmitted plastic cuvette.
    Cuvette Washing 5 channels automatic washing system for reaction cuvettes
    Probe and Mixer Washing Independent washing station for each sample/reagent probe and mixer
    Alarm Automatic alarm for malfunction (eg: collision, reagent/sample insufficiency, full waste container, etc)
    Calibration Linear, Non-linear, Multi standard, K factor
    Sample Dilution/Retest Sample can be diluted and retested if results are out of range or sample is insufficient. Also with calibrator auto diluting function
    Wavelength 340-810nm
    Light Source Halogen lamp
    Absorbance Resolution 0.0001A
    Absorbance Accuracy ±0.0003A(0-2.5A)
    Repeatability ±2%
    Quality Control Levy jennings quality control program at 3 levels
    Barcode Optional
    Printer Multiple report formats available
    System Windows XP; Windows 7
    Power Supply 220V or 110V, 50~60Hz
    Dimension 65×53×52(cm)
    Weight 45kg

     




     
     