Categories

YSAV610plus medical anesthesia machine System ventilator

Simplicity: easy to use, easy to move with 4 wheels.
Patient Centered Ventilation: Precision in an anesthesia ventilator, from conventional ventilation to advanced modes, including 5 modes: IPPV; A/C; PCV; SIMV; SIGH, MAUN.
Designed and manufactured by 
  • Brandysenmed
  • Model numberYSAV610plus
  • Warranty1 year
  • Customizedsupport
  • Lead time7days
  • Packagewooden
  • Description
    YSAV610plus medical anesthesia machine System ventilator
    SSpecifications       
    Ventilation modes IPPV; A/C; PCV; SIMV; SIGH; MAUN
     
    Ventilator parameter ranges
    Tidal volume(Vt) 0, 20 mL ~ 1500 mL
    Frequence(Freq) 1bpm ~ 100 bpm
    I/E 4:1~1:8
    PEEP 0cmH2O ~ 30 cmH2O
     
    Monitored Parameters
    Frequence(Freq) 0 /min ~ 100 /min
    Tidal volume(Vt) 0 mL ~ 2000 mL
    MV 0 L/min ~ 100 L/min
    Oxygen concentration 15 % ~ 100 %
     
    Oscillographs display 
    P-T(pressure – time)  
    F-T(flow – time)  
    P-V loop (pressure – volume loop)  
    V-T(volume – time )  
     
    Size
    Wooden case packing size:L 870*W 890*H 1600mm , GW:190KG ;
    Anesthesia Machine size:L930*W750*H 1405 mm ,NW:124KG
     
    Alarm and protection  
    The AC power failure alarm Power failure or no connection
    Internal battery backup low voltage alarm <11.3±0.3V
    No tidal volume ≤5Ml within 6s
    High oxygen concentration alarm
    Low oxygen concentration alarm
    19%-100%
    18%-99%
    High Airway  pressure alarm
    Low Airway  pressure alarm
    High Minute Volume alarm
    Low Minute Volume alarm
    Continuous Pressure alarm
    20cmH2O-100cmH2O
    0cmH2O-20cmH2O
    Adult(5L/min-20L/min) Paed(1L/min-15L/min
    0-10L/min)
    (PEEP+1.5kPa) over 16s
    Suffocation warning 5s-60s no spontaneous ventilation
    The maximum limited pressure   <12.5 kPa
    Fan error
    Oxygen deficit
    Show on screen
    Show on screen
     
    Working conditions
     
    Gas source O2,N2O,AIR
    Pressure 280kPa-600kPa
    Voltage 100-240V
    Power frequency 50/60Hz
    Input power 80VA
     