1:
|
Summary of main specifications and system
|
|
1.1
|
Portable type full digital color Doppler ultrasound main unit
|
|
1.2
|
Ultrasonic host operating system: Windows operating system
|
|
1.3
|
Applications:
Abdomen, obstetrics, gynecology, heart, urinary system, small organs, superficial, blood vessels, pediatrics, newborns, musculoskeletal |
|
1.4
|
Probes:
Convex probe, Tran-vaginal probe, Linear probe, Micro-convex probe, Cardiac probe,4D Volume probe |
|
1.5
|
Applications and report:
Abdominal,OB,GYN,Cardiac,Urinary,Small Parts,Superficial, Vascular, Pediatrics, Advanced measurement software packages, report software packages, case management software packages, etc. |
|
☆1.6
|
carotid artery intima measurement thickness(IMT)
|
|
☆1.7
|
Automatic spectral envelope measurement
|
|
1.8
|
Full digital transmission and reception of beam synthesizer
|
|
1.9
|
Color Doppler imaging(C)
|
|
1.10
|
Pulse Doppler Imaging(PW)
|
|
|
|
|
1.12
|
Continuous wave Doppler imaging(CW)
|
|
☆1.13
|
B/C/D Real-time three synchronous imaging
|
|
☆1.14
|
Power Doppler imaging(PDI)
|
|
☆1.15
|
Direct power Doppler imaging(DPDI)
|
|
1.16
|
M mode imaging
|
|
☆1.17
|
Anatomic M mode imaging
|
|
☆1.18
|
Color Doppler M mode imaging
|
|
|
|
|
☆1.20
|
Tissue Doppler imaging(TDI)
|
|
☆1.21
|
Strain rate imaging (SRI)
|
|
1.22
|
Tissue harmonic imaging(THI)
|
|
1.23
|
Fusion harmonic imaging(FHI)
|
|
1.24
|
Speckle Reduce imaging(SRI)
|
|
|
|
|
☆1.26
|
Deflection imaging
|
|
☆1.27
|
Trapezoidal imaging
|
|
1.28
|
Adaptive velocity optimization
|
|
☆1.29
|
Free hand 3D
|
|
1.30
|
Real time 3D imaging(3D/4D)
|
|
1.31
|
DICOM3.0
|
|
1.32
|
Monitor:15 inch,high definition ultrasonic display
|
|
|
|
|
1.34
|
Physical clipboard: save the image on the left side of the screen, which can be directly saved or deleted.
|
|
1.35
|
The system has the function of on-the-spot upgrade
|
|
1.36
|
Presupposition: for different inspection of the viscera, preset the inspection conditions for the best image, reduce the adjustment of the operation, and the commonly used external adjustment and combination regulation.
|
|
1.37
|
Probe interface: 3 active
|
|
1.38
|
10 kinds of system language
|
|
1.39
|
Depth:≥360mm;
|
|
1.40
|
Extended imaging
|
|
2:
|
Probes(more probes options on additional file)
|
|
2.1 Convex probe
|
Fundamental Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz/3.5MHz/4.0MHz/4.6MHz/5.0MHz/5.4MHz,
Harmonic Frequency: 4.0MHz/4.6MHz/5.0MHz, |
|
2.2
Linear probe |
Fundamental Frequency:4.0MHz/4.6MHz/5.0MHz/6.0MHz/7.0MHz/8.0MHz/9.2MHz/10.0MHz/12.0MHz/13.3MHz,
Harmonic Frequency: 8.0MHz/9.2MHz/10.0MHz, |
|
2.3 Trans-vaginal probe
|
Fundamental Frequency:
3.0MHz/3.5MHz/4.0MHz/5.0MHz/5.4MHz/6.0MHz/7.0MHz/8.0MHz/10.0MHz, Harmonic Frequency: 6.0MHz/7.0MHz/8.0MHz, |
|
2.4 Micro-convex probe
|
Fundamental Frequency:
3.0MHz/3.5MHz/4.0MHz/5.0MHz/5.4MHz/6.0MHz/7.0MHz/8.0MHz, Harmonic Frequency: 6.0MHz/7.0MHz/8.0MHz, |
|
2.5 Cardiac probe
|
Fundamental Frequency:1.7MHz/1.9MHz/2.1MHz/2.5MHz/3.0MHz/3.4MHz/3.8MHz/4.2MHz/5.0MHz,
Harmonic Frequency: 3.4MHz/3.8MHz/4.2MHz, |
|
2.6
4D Volume probe |
Fundamental Frequency:
2.0MHz/2.5MHz/3.0MHz/3.3MHz/3.7MHz/4.0MHz/5.0MHz/6.0MHz, Harmonic Frequency: 4.0MHz/5.0MHz/6.0MHz, |
|
3:
|
2D imaging mode
|
|
3.1
|
Gain:0-100,Step 2 adjustable
|
|
3.2
|
TGC:8 segment adjustable
|
|
3.3
|
Maximum focus point:≥7, which can be moved throughout the whole process.
|
|
3.4
|
Speckle reduction:0-5,5 level
|
|
3.5
|
Space Synthesis:0-2,2 level(Liner probe: 3 level, cardiac probe:0)
|
|
3.6
|
Dynamic:30-180,35 level,step 5 adjustable
|
|
3.7
|
Line density:low、middle、high,3 level
|
|
3.8
|
Frame correlation:0-4,4 level
|
|
3.9
|
Noise reduction:0-5,5 level
|
|
3.10
|
Edge Enhancement:0-5,5 level
|
|
3.11
|
Sound power:2-10, 9 level
|
|
3.12
|
Grey scale:0-67, 67 level
|
|
3.13
|
False color:0-67,67 level
|
|
3.14
|
Image style:Soft-Comparison,2 level
|
|
|
The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted
|
|
4:
|
Color Doppler imaging mode
|
|
4.1
|
Blood gain:0-100,Step 2
|
|
4.2
|
Parameter display:Velocity、Variance
|
|
4.3
|
B-Restrain(B/W restrain):0-7, 7 level
|
|
4.4
|
Speed Through:0-8, 8 level
|
|
4.5
|
Sampling number:6-24, 7 level
|
|
4.6
|
Blood flow preferred:0-8, 8 level
|
|
4.7
|
Filtering:1-6, 6 level
|
|
4.8
|
Sound power:2-6, 4 level
|
|
4.9
|
Noise reduction:0-4, 4 level
|
|
4.10
|
Smooth treatment:0-4, 4 level
|
|
4.11
|
Frame correlation:0-6, 6 level
|
|
4.12
|
Chromatography(Blood flow graph):0-37, 37 level
|
|
4.13
|
Line density:Low-Middle-High, 3 level
|
|
4.14
|
Frequency:4 level adjustable
|
|
4.15
|
Velocity:Minimum 0.4K,Maximum 40.5K
Convex probe:0.4K-4.3K-38.5K Linear probe:0.4K-14.7K-39.0K Trans-vaginal probe:0.4K-7.8K-39.7K Volume probe:0.4K-4.2K-34.8K Micro-convex probe:0.4K-10.3K-40.5K cardiac probe:0.4K-7.8K-39.7K |
|
|
PS:The frequency of the probe changes and the frequency value changes
|
|
|
PS:Frame rate changes with speed
|
|
5:
|
Pulse wave Doppler(PW)
|
|
1
|
Gain:0-100,Step 2
|
|
5.2
|
Spectrum envelope function: real time automatic spectrum envelope, manual spectrum envelope, and other modes. The system automatically analyses and displays various data such as PSV, EDV, RI, PI, S/D, ACC, HR and so on. Can wake up or close
|
|
5.3
|
Sample volume:0.5mm~30mm
|
|
5.4
|
Blood angel:-75—75 degree,Step 5
|
|
5.5
|
False color:0-67, 67 level
|
|
5.6
|
Dynamic range:20-40, 4 level
|
|
5.7
|
Filter:0-9, 9 level
|
|
5.8
|
Smooth treatment:1-4, 4 level
|
|
5.9
|
Sound power:2-5, 4 level
|
|
5.10
|
Volume:0-100, 10 level,Step 10
|
|
5.11
|
Audio filtering:0-4, 4 level
|
|
5.12
|
Base line:-1.0~1.0,
|
|
5.13
|
Grey map:0-67, 67 level
|
|
5.14
|
Scan velocity:100-500, 6 level
|
|
5.15
|
PRF:Minimum 0.5K,Maximum 87.5K
Convex probe:0.5K-4.3K-63.3K Linear probe:0.5K-14.5K-78.4K Trans-vaginal probe:0.5K-8.1K-78.4K Volume probe:0.5K-4.2K-53.8K Micro-convex probe:0.5K-10.3K-81.1K cardiac probe:0.5K-4.3K-87.5K |
|
5.16
|
Frequency:4 level
|
|
|
PS:The frequency of the probe changes and the PRF value changes
|
|
|
PS:The frequency of the probe changes and the frequency value changes
|
|
6:
|
Continuous Wave Doppler(CW)
|
|
6.1
|
Support probe:Cardiac probe
|
|
6.2
|
Adjustment of B mode parameters is switchable
|
|
6.3
|
Gain:0-100,Step 2
|
|
6.4
|
Sampling line position is adjustable
|
|
6.5
|
PRF:0.9K~36.1K
|
|
6.6
|
Baseline:-1.0~1.0
|
|
6.7
|
Blood angel:-75~75 degree
|
|
6.8
|
Grey map:0-67
|
|
6.9
|
Scan velocity:100-300
|
|
6.10
|
False color:0-67
|
|
6.11
|
Dynamic range:20-40
|
|
6.12
|
Filtering:0-9,9 level
|
|
6.13
|
Smooth treatment:1-4
|
|
6.14
|
Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz,4 level adjustable
|
|
6.15
|
Sound power:2-5
|
|
6.16
|
Volume:0-100
|
|
6.17
|
Audio Filtering:0-4
|
|
☆7:
|
Anatomical M imaging
|
|
7.1
|
Support probe:Convex probe, Linear probe,Cardiac probe
|
|
7.2
|
Adjustment of B mode parameters is switchable
|
|
7.3
|
Gain:0-100,Step 2
|
|
7.4
|
M Sampling line angel is adjustable
|
|
7.5
|
M Sampling line length is adjustable
|
|
7.6
|
Sampling line:3,Can be displayed or hidden separately
|
|
☆8:
|
Blood flow M model(MC)
|
|
8.1
|
Adjustment of B mode parameters is switchable
|
|
8.2
|
Gain:0-100,Step2
|
|
8.3
|
MC Sampling line angel is adjustable
|
|
8.4
|
MC Sampling line length is adjustable
|
|
8.5
|
Frequency:4 level
|
|
8.6
|
Sampling number:6-24
|
|
8.7
|
Speed through:0-8, 8 level
|
|
8.8
|
Scan velocity:150-500
|
|
8.9
|
Frame correlation:0-6, 6 level
|
|
8.10
|
Filtering:1-6,6 level
|
|
8.11
|
Blood flow preferred:0-8,8 level
|
|
8.12
|
Smooth treatment:0-4,4 level
|
|
8.13
|
Map:0-37, 37 level
|
|
☆9:
|
Elastography
|
|
9.1
|
Adjustment of B mode parameters is switchable
|
|
9.2
|
Gain:0-100,Step 2
|
|
9.3
|
B/E,Double real-time display on the same screen
|
|
9.4
|
Probe displacement curve display:Up/Down
|
|
9.5
|
Pressure indicator bar display
|
|
9.6
|
Frequency:8-9 level,Adjustable;According to the probe display
|
|
9.7
|
Noise reduction:0-2, 2 level
|
|
9.8
|
Frame correlation:0-3, 3 level
|
|
9.9
|
Comparison:0-13, 13 level
|
|
9.10
|
False color:0-3, 3 level
|
|
9.11
|
Don’t support cardiac probe
|
|
☆10:
|
Tissue Doppler imaging(TDI)
|
|
10.1
|
Support probe: Cardiac probe
|
|
10.2
|
Adjustment of B mode parameters is switchable
|
|
10.3
|
Gain:0-100,step 2
|
|
10.4
|
ROI area adjustable
|
|
10.5
|
Sampling number:6-24
|
|
10.6
|
Velocity:0.4K-8.0K
|
|
10.7
|
Frame correlation:0-6,6级
|
|
10.8
|
Tissue preferred:0-7, 7 level
|
|
10.9
|
Frequency:2.0MHz/2.3MHz/2.5MHz/3.0MHz
|
|
10.10
|
Support color reversal
|
|
☆11:
|
Strain rate imaging
|
|
11.1
|
Support probe: Cardiac probe
|
|
11.2
|
Adjustment of B mode parameters is switchable
|
|
11.3
|
ROI area adjustable
|
|
11.4
|
Gain:0-100,Step 2
|
|
11.5
|
Sampling number:6-24,6 level
|
|
11.6
|
Axial average:1-4, 4 level
|
|
11.7
|
Velocity:0.4K-8K
|
|
11.8
|
Frame correlation:0-6, 6 level
|
|
11.9
|
Tissue optimization:0-7,7 level
|
|
☆12:
|
Panoramic imaging
|
|
12.1
|
Support probe:Linear probe
|
|
12.2
|
Speckle Reduction:0-5, 5 level
|
|
☆13:
|
Deflection imaging
|
|
13.1
|
Support probe:Linear probe
|
|
13.2
|
Adjustment of B mode parameters is switchable
|
|
13.3
|
Deflection angel:8 level
|
|
13.4
|
Speckle reduction:0-5, 5 level
|
|
13.5
|
Dynamic rate:30-180,Step 5
|
|
13.6
|
Line density:low-middle-high,3 level
|
|
13.7
|
Frame Correlation:0-4, 4 level
|
|
13.8
|
False color:0-67, 67 level
|
|
13.9
|
Image style:Soft-Comparison,2 level
|
|
13.10
|
Noise reduction:0-5, 5 level
|
|
13.11
|
Edge Enhancement:0-5,5 level
|
|
13.12
|
Sound power:2-10,8 level
|
|
13.13
|
Grey map:0-67,67 level
|
|
☆14:
|
Trapezoidal imaging
|
|
14.1
|
Probe support:linear probe
|
|
14.2
|
Adjustment of B mode parameters is switchable
|
|
14.3
|
Deflection angel:8 level
|
|
14.4
|
Speckle reduction:0-5, 5 level
|
|
14.5
|
Dynamic rate:30-180,Step 5
|
|
14.6
|
Line density:low-middle-high,3 level
|
|
14.7
|
Frame Correlation:0-4, 4 level
|
|
14.8
|
False color:0-67, 67 level
|
|
14.9
|
Image style:Soft-Comparison,2 level
|
|
14.10
|
Noise reduction:0-5, 5 level
|
|
14.11
|
Edge Enhancement:0-5,5 level
|
|
14.12
|
Sound power:2-10,8 level
|
|
14.13
|
Grey map:0-67, 67 level
|
|
14.14
|
Space Synthesis:0-2, 2 level
|
|
☆15
|
Freehand 3D imaging
|
|
15.1
|
Support probe:convex probe, linear probe
|
|
15.2
|
Display model: 4 pictures
|
|
15.3
|
Image Rotation X/Y/Z Axis
|
|
15.4
|
Multi-slice Visibility
|
|
16
|
Real-time 4D imaging
|
|
16.1
|
Support probe: 4D volume probe
|
|
16.2
|
Adjustment of B mode parameters is switchable
|
|
16.3
|
Gain:0-100,Step 2
|
|
16.4
|
Display model:one image、two images、four images
|
|
16.5
|
Image Rotation:X/Y/Z Axis
|
|
16.6
|
Multi-slice Visibility
|
|
16.7
|
Light&Shade inversion
|
|
16.8
|
Smooth:0-4, 4 level
|
|
16.9
|
Threshold level:0-129, Step 3
|
|
16.10
|
Transparency:1-509,Step 10
|
|
16.11
|
Render type:4 kinds,Surface、maximum、minimum、perspective
|
|
17:
|
Extended Imaging
|
|
17.1
|
Gain:0-100,Step 2
|
|
17.2
|
TGC:8 segment adjustable
|
|
17.3
|
Maximum focus point:≥7, which can be moved throughout the whole process.
|
|
17.4
|
Speckle reduction:0-5,5 level
|
|
17.5
|
Space Compound:0-2,2 level(Linear probe: 3 level,don’t support cardiac probe)
|
|
17.6
|
Dynamic range:30-180,35 level,Step 5
|
|
17.7
|
Line density:Low、Middle、High,3 level
|
|
17.8
|
Frame correlation:0-4,4 level
|
|
17.9
|
Noise reduction:0-5,5 level
|
|
17.10
|
Edge enhancement:0-5,5 level
|
|
17.11
|
Sound power:2-10, 9 level
|
|
17.12
|
Grey map:0-67, 67 level
|
|
17.13
|
False color:0-67,67 level
|
|
17.14
|
Image style:Soft-Comparison,2 level
|
|
17.15
|
Extended level: Maximum 72 level
Convex probe:9 level Trans-vaginal probe:72 level Micro-convex probe:29 level Cardiac probe:40 level 4D Volume probe:17 level |
|
|
PS:The screen has real-time display of voice power, probe frequency, dynamic range, pseudo color, gray scale and other 11 parameters can be adjusted
|
|
|
PS: When the probe scan range reaches the maximum, the space synthesized is 0.
|
|
18:
|
Measurement and analysis function:
|
|
18.1
|
General measurement: Distance, area, ellipse, cross line, angle, distance ratio, volume, Volume (ellipse), area ratio, diameter, joint angle
|
|
18.2
|
Cardiac:Automatic spectrum envelope、LV、Main Pulmonary artery diameter、RVEDd、RVEDs、LVM、LAV、HR、MVF、AO、AR、LVOT、TVF、Pulmonic valve、Pulmonary vein、RV、Doppler fetal heart sound、LVET、LVM、LVMI、AV
|
|
18.3
|
Vascular: carotid intima (IMT), length stenosis ratio, area stenosis ratio, IMT (back wall), IMT (front wall)
|
|
18.4
|
OB:Fetal routine、AFI、TW、GS、CRl、OFD、HL、ulna、NT、Fibula、Nbonel、Radial、Tibia
|
|
18.5
|
GYN:uterus、cervix、corpus uteri/cervix uterus、left ovarian vein、right ovarian vein、dominant follicle、intima thickness
|
|
18.6
|
Urology: prostate、residual urine、left kidney、right kidney、left suprarenal vein、right suprarenal、left testis、right testis、left seminal vesicle、right seminal vesicle
|
|
18.7
|
Abdomen:liver、CHD、partal vein diameter、cholecyst、CBD、pancreas、spleen、Internal diameter of abdominal aorta、kidney
|
|
18.8
|
Small parts:Thyroid
|
|
18.9
|
Software package:Measurement package、Software package、Medical records management software package
|
|
19:
|
Graphic and text management system
|
|
19.1
|
Host build in 2 hard disk (SSD 256GB), Start fast and stable
|
|
19.2
|
Movie playback:≥1200 frames
|
|
19.3
|
Internal file information management system: can record patient number, name, check number, check date and so on, and can be searched and managed by numbering, checking number, name and so on.
|
|
19.4
|
Type of report is 16
|
|
19.5
|
One key fast report graphic and text management
|
|
20:
|
Interface
|
|
20.1
|
USB interface: 2
|
|
20.2
|
HDMI interface: 1
|
|
20.3
|
VGA interface:1
|
|
|
|
|
|
|
|
21:
|
Configuration
|
|
21.1
|
Portable full digital color Doppler ultrasound diagnostic system main unit
|
|
21.2
|
Probe options: convex, linear , Trans-vaginal , phased array, micro-convex, 4D volume and trans-rectal
|
|
|
|
|
22:
|
Technology, after-sales service and other requirements
|
|
22.1
|
Two years warranty for both main unit and probes
|
|
22.2
|
Manufacturer has ISO13485 certification and EU CE certification.
|